5M EMC Chamber
    Publish time 2016-09-10 20:02    
5M EMC chamber is on the basis of electromagnetic shielding room, with the walls and ceilings mounted on the electromagnetic wave absorption material and the ground is the ideal reflection surface so as to simulate the open site test conditions.
5M EMC Chamber

5M EMC Chamber

Performance requirements
1, shielding performance

Field Frequency  Attenuation
Magnetic field 10kHz 〉70dB

100kHz     〉90dB

1MHz  〉110dB
Electric field 100kHz-100MHz  100dB
Plane wave 100MHz-1GHz 〉100dB
Microwave  1GHz-40GHz   〉100d

2Quiet Zone Normalized Site Attenuation (NSA):5m test distance, 3m dead zone (3m diameter, 2m high) and 3m test distance, 2m dead zone (2m diameter, 2m high) case test. From 30 MHz to 1 GHz in accordance with ANSI C63.4-2003 standard test, the actual NSA and the theoretical NSA compared to no more than ± 3.5 dB.
 
3Quiet zone transmission loss: 1GHz-18GHz using transmission loss (TL) test method to test, in the 3m test distance, 2m dead zone (2m diameter, 2m high) case test; TL value is not greater than ± 3.5dB.

4, the site voltage standing wave ratio: from 1GHz - 18 GHz in accordance with CISPR16-1-4 ED2 2006 and CISPR / A / 710 FDIS, diameter 3m, 2m high cylinder static zone, the test distance of 5m, two height, two Polarization direction field voltage standing wave ratio is better than 6dB.
  5, the field uniformity: in accordance with the latest version of IEC 61000-4-3 test, from 26MHz-18GHz frequency range of 16 test points in the 12 points of the uniformity of 0 ~ +6 dB, the standard field for the turntable 0.8m-2.3m in the range of 1.5mx 1.5m vertical plane.
  6, the background noise: dark room space environment background noise (dark room turntable, antenna lift tower, lights and TV monitors and other equipment to open the state) test level than in the 30MHz ~ 1GHz frequency range EN55022, GB9254 standard The prescribed limit.